3 leg isolated rtd input (/n2), Extended input type (/n3) – Yokogawa Removable Chassis DX1000N User Manual

Page 308

Advertising
background image

12-14

IM 04L41B01-01E

3 Leg Isolated RTD Input (/N2)

Item

Specifications

Input terminal

All the RTD input terminals (A, B, and b) are isolated on each channel.

Applies to DX1006, DX1012, DX1006N, and DX1012N

Note: On the DX2004 and DX2008 standard models, all the terminals (A, B, and b) are already

isolated on each channel

Extended Input Type (/N3)

Item

Specifications

Measurement/display accuracy

Under standard operating conditions

Kp vs Au7Fe

0.0 to 300.0K

PLATINEL

0.0 to 1400.0°C

PR40-20

0.0 to 1900.0°C

NiNiMo

0.0 to 1310.0°C

W/WRe26

0.0 to 2400.0°C

Type N(AWG14) 0.0 to 1300.0°C
XK GOST

*2

–200.0 to 600.0

Pt50

–200.0 to 550.0°C

Ni100(SAMA)

–200.0 to 250.0°C

Ni100(DIN)

–60.0 to 180.0°C

Ni120

–70.0 to 200.0°C

J263*B

0.0 to 300.0K

Cu53

–50.0 to 150.0°C

Cu100

–50.0 to 150.0°C

Pt25

–200.0 to 550.0°C

Pt100GOST

*2

–200.0 to 600.0°C

Cu100 GOST

*2

–200.0 to 200.0°C

Cu50 GOST

*2

–200.0 to 200.0°C

Cu10 GOST

*2

–200.0 to 200.0°C

Pt46 GOST

*2

–200.0 to 550.0°C

Pt200 (WEED)

*3

–100.0 to 450.0°C

*1 Measuring current i = 1 mA
*2 Available for release numbers 3 and later.

Measurement

Range

Measurement Accuracy

Max.

Resolution

Input Type

Thermocouple

RTD

*1

0 to 20 K

Within ±4.5 K

20 to 300 K

Within ±2.5 K

±(0.25% of rdg + 2.3°C)
0 to 450°C

Accuracy not guaranteed

450 to 750°C

±(0.9% of rdg + 3.2°C)

750 to 1100°C

±(0.9% of rdg + 1.3°C)

1100 to 1900°C ±(0.9% of rdg + 0.4°C)
±(0.25% of rdg + 0.7°C)
0 to 400°C

±15.0°C

400 to 2400°C

±(0.2% of rdg + 2.0°C)

±(0.2% of rdg + 1.3°C)
–200 to –100°C ±(0.25% of rdg +1.0°C)
–100 to 600°C

±(0.25% of rdg + 0.8°C)

±(0.3% of rdg + 0.6°C)
±(0.15% of rdg + 0.4°C)
±(0.15% of rdg + 0.4°C)
±(0.15% of rdg + 0.4°C)
0 to 40 K

Within ±3.0 K

40 to 300 K

Within ±1.0 K

±(0.15% of rdg + 0.8°C)
±(0.2% of rdg + 1.0°C)
±(0.15% of rdg + 0.6°C)
±(0.15% of rdg + 0.3°C)
±(0.15% of rdg + 0.3°C)
±(0.4% of rdg + 0.5°C)
±(1.5% of rdg + 3.0°C)
±(0.3% of rdg + 0.8°C)
±(0.3% of rdg + 0.6°C)

Within ±13.5 K
Within ±7.5 K
±(0.25% of rdg + 8.0°C)
Accuracy not guaranteed
±(0.9% of rdg + 15.0°C)
±(0.9% of rdg + 6.0°C)
±(0.9% of rdg + 3.0°C)
±(0.5% of rdg + 3.5°C)
±30.0°C
±(0.4% of rdg + 4.0°C)
±(0.5% of rdg + 7.0°C)
±(0.5%of rdg +5.0°C)
±(0.5%of rdg +4.0°C)
±(0.6% of rdg + 3.0°C)
±(0.3% of rdg + 2.0°C)
±(0.3% of rdg + 2.0°C)
±(0.3% of rdg + 2.0°C)
Within ±9.0 K
Within ±3.0 K
±(0.3% of rdg + 4.0°C)
±(0.4% of rdg + 5.0°C)
±(0.3% of rdg + 3.0°C)
±(0.3% of rdg + 1.5°C)
±(0.3% of rdg + 1.5°C)
±(0.8% of rdg + 2.5°C)
±(3.0% of rdg + 15.0°C)
±(0.6% of rdg + 4.0°C)
±(0.6% of rdg + 3.0°C)

0.1K

0.1°C

0.1K

0.1°C

A/D integration time: 16.7 ms or more

A/D integration time:

1.67 ms

*

3 Available for release numbers 4 and later. Double the resistance of a 100 ohm Platinum (TCR = .003902 ohms/ohm/°C) Curve A resistor

made by Weed Instrument.

Input source resistance

Thermocouple input: 2 kΩ or less

RTD input:

1 Ω or less per wire (The resistance of all three wires must be equal).

Ambient temperature influence (applies when the A/D integration time is 16.67 ms or greater)
TC input

±(0.1% of rdg + 0.05% of range) or less, excluding the error of reference junction compensation

RTD input

±(0.2% of range + 2 digits) or less

Input source resistance
TC input

With variation of +1 kΩ: ±10 µV or less

RTD input

With variation of 1 Ω per wire (resistance of all three wires must be equal): ±(0.1% of rdg + 1 digit)

or less

With maximum difference of 100 mΩ between wires: Approx. 1 °C

12.5 Options

Advertising
This manual is related to the following products: