12-3

M-4660

Troubleshooting and Maintenance

1

2

3

4

5

6

7

8

9

10

11

12

13

App

Index

Code

Message

Explanation/Countermeasures/Ref. section

102

SUBNET mask is incorrect.

Check the subnet mask.
Section 2.2 in the communication manual

103

The net part of default gateway is not equal to that of IP
address.

Check the IP address.
Section 2.2 in the communication manual

105

This port number is already in use. Please enter a different
number.

Enter a different port number for each function.
Section 7.1 in the communication manual

113

Password entered is incorrect.

Enter the correct password.
See section 8.3

119

This user name is unable to use this mode.

A user-level user cannot enter the Basic Setting
mode.
See section 8.2

120

Measured value is incorrect. (in ascending order)

Set the calibration correction value to a value greater
than the previous value.
See section 3.9

122

Measured value exceeds the range setting.

Check the channel input or the channel range setting.
See sections 3.3 and 3.9

125

Character entry cannot be performed.

The MV is not showing a display used to enter
character strings.
See section 9.7

126

You cannot use the same password.

Specify a different password.
See section 8.3

127

Report kind overlaps and cannot be set up.

Change the overlapped report data type.
See section 10.5

129

IP address is not set.

Set the MV IP address.
Section 2.2 in the communication manual

131

You have exceeded the available channel capacity.

You cannot connect more than 240 channels.
Section 2.9 in the communication manual

132

You have exceeded the available number of commands.

The maximum number of commands that can be
sent is 16. The modules that can be set with a
single command are consecutive modules that can
be automatically set. Change the MW100 module
configuration so that there are no empty slots.
Section 2.9 in the communication manual

133

External I/O auto setting information is not available.

Below are the possible causes. Check them.
• The MW100 is in calibration mode. Change to the

setting mode or measurement mode.

• The measurement module may not have been

detected. Perform system reconfiguration.

• There are no modules that can be automatically set.

Check the modules.

• An IP address has not been assigned to the

MW100. Set the IP address.

• The Modbus server of the MW100 is turned OFF.

Turn the server ON.

Section 2.9 in the communication manual

134

Auto setting has already been executed.

You cannot set an MW100 that has been
automatically set.
Section 2.9 in the communication manual

135

External I/O cannot be found.

Check the Ethernet connection.
Section 2.2 in the communication manual

136

External I/O start cannot be executed.

The current MW100 settings do not allow the
measurement to be started. Check the settings.

137

DNS for this device is not set.

Set the MV DNS parameter.
Section 2.2 in the communication manual

12.1 List of Messages