3-20

M-4660

3.11 Setting Range-Out Detection for Linearly

Scaled Measurement Channels

Display

RD-MV1000

Press MENU > then select Menu tab > Basic setting mode > Environment tab > Input,
Alarm

RD-MV2000

Press MENU and then select Menu tab > Basic setting mode > Environment tab >
View, Message, Input, Alarm.

Settings

• Input > Value on over-range

Setting

Description

Free

Ignoring the decimal point, less than –30000 is a negative range-out, and above
30000 is a positive range-out. They are displayed as –Over and +Over.

Over

A value less than –5% of the scale is a negative range-out, and a value greater than
105% of the scale is a positive range-out. They are displayed as –Over and +Over.
Example: If the scale is 0.0 to 200.0, a value less than –10.0 is a negative range-

out, and a value greater than 210.0 is a positive range-out.

Note

With computation and report functions such as TLOG and CLOG, you can specify how the MV
will handle scale range out values.

See section 10.1.