Three-wire isolated rtd input (/n2), Extended input (/n3), Remote control (/r1) – Yokogawa Button Operated MV2000 User Manual

Page 274

Advertising
background image

13-12

M-4660

Three-Wire Isolated RTD Input (/N2)

Item

Specifications

Input terminal

Isolation on each channel
Applies to RD-MV1006, RD-MV1012, RD-MV1024, RD-MV2010, RD-MV2020, RD-MV2030, RD-
MV2040, and RD-MV2048

Extended Input (/N3)

Item

Specifications

Measurement/display accuracy

Under standard operating conditions

Kp vs Au7Fe

0.0 to 300.0 K

PLATINEL

0.0 to 1400.0°C

PR40-20

0.0 to 1900.0°C

NiNiMo

0.0 to 1310.0°C

W/WRe26

0.0 to 2400.0°C

Type N (AWG14) 0.0 to 1300.0°C
Pt50

–200.0 to 550.0°C

Ni100 (SAMA)

–200.0 to 250.0°C

Ni100 (DIN)

–60.0 to 180.0°C

Ni120

–70.0 to 200.0°C

J263*B

0.0 to 300.0 K

Cu53

–50.0 to 150.0°C

Cu100

–50.0 to 150.0°C

Pt25

–200.0 to 550.0°C

32 to 2552°F
32 to 3452°F

32 to 2390°F
32 to 4352°F

32 to 2372°F
–328.0 to 1112.0°F
–328.0 to 482.0°F
–76.0 to 356.0°F
–94.0 to 392.0°F

–58.0 to 302.0°F
–58.0 to 302.0°F
–328.0 to 1022.0°F

* Measuring current i = 1 mA

Measurement Range

Measurement Accuracy

Max.

Reso-

lution

Input Type

TC

RTD*

0 to 20 K

Within 4.5 K

20 to 300 K

Within 2.5 K

(0.25% of rdg + 2.3°C)
0 to 450°C

Accuracy not guaranteed

450 to 750°C

(0.9% of rdg + 3.2°C)

750 to 1100°C

(0.9% of rdg + 1.3°C)

1100 to 1900°C (0.9% of rdg + 0.4°C)
(0.25% of rdg + 0.7°C)
0 to 400°C

15.0°C

400 to 2400°C

(0.2% of rdg + 2.0°C)

(0.2% of rdg + 1.3°C)
(0.3% of rdg + 0.6°C)
(0.15% of rdg + 0.4°C)
(0.15% of rdg + 0.4°C)
(0.15% of rdg + 0.4°C)
0 to 40 K

Within 3.0 K

40 to 300 K

Within 1.0 K

(0.15% of rdg + 0.8°C)
(0.2% of rdg + 1.0°C)
(0.15% of rdg + 0.6°C)

Within 13.5 K
Within 7.5 K
(0.25% of rdg + 8.0°C)

Accuracy not guaranteed

(0.9% of rdg + 15.0°C)
(0.9% of rdg + 6.0°C)
(0.9% of rdg + 3.0°C)
(0.5% of rdg + 3.5°C)
30.0°C
(0.4% of rdg + 4.0°C)
(0.5% of rdg + 7.0°C)
(0.6% of rdg + 3.0°C)
(0.3% of rdg + 2.0°C)
(0.3% of rdg + 2.0°C)
(0.3% of rdg + 2.0°C)
Within 9.0 K
Within 3.0 K
(0.3% of rdg + 4.0°C)
(0.4% of rdg + 5.0°C)
(0.3% of rdg + 3.0°C)

0.1 K

0.1°C

0.1 K

0.1°C

A/D integration time:

16.7 ms or more

A/D integration time:

1.67 ms

Input source resistance

Thermocouple input: 2 kΩ or less
RTD input: 1 Ω or less per wire (The resistance of all three wires must be equal)

Ambient temperature influence (applies when the A/D integration time is 16.67 ms or greater)

TC input

±(0.1% of rdg + 0.05% of range) or less, excluding the reference junction compensation error

RTD input

±(0.2% of range + 2 digits) or less

Signal source resistance

TC input

With a variation of signal source resistance +1 kΩ: ±10 µV or less

RTD input

With a variation of 1 Ω per wire (same resistance for all three wires): ±(0.1% of rdg + 1 digit) or
less
With a maximum difference of 100 mΩ between wires: Approx. 1°C

Remote Control (/R1)

Item

Specifications

Number of input terminals 8
Input type

Isolated from the main circuitry through a photocoupler, built-in isolated power supply for the input
terminals, and shared common.

Input type and signal level

Voltage-free contact

Contact closed at 200 Ω or less and contact open at 100 kΩ or greater.

Open collector

ON voltage: 0.5 V or less (sink current 30 mA or more), leakage current when OFF: 0.25 mA or
less

Allowable input voltage

5 VDC

Signal type

Level or edge (250 ms or more)

Action

Executes a specified action by applying a given signal to the remote signal input terminal
Action assignment: Set using the event action function

13.5 Options

Advertising