Appendix 1 Data File Size

This section explains how to calculate the file sizes of display data files and event data
files. The calculation examples are for when the MV is set to record only display data or
only event data. If you are recording both display and event data, calculate the data size
of each and add them together.
Use the calculated file size as a rough estimate.

File Size

A file consists of the following:
sampled-data related information + sampled data

The maximum size of a single display data file or event data file is 8 MB.

Sampled-Data Related Information Size

Item

Size in Bytes

File header

216

Channel information

88×N + 32

Group information

96Ч10 + 32 = 992
96Ч36 + 32 = 3,488

(RD-MV1000)
(RD-MV2000)

Message information

104×50 + 32(an add message area is reserved by default)

Batch information

832

Sampled data header

80 + 32 + N×8 + 16 + 2

Alarm information header

24 + 8 (add the size of this item even if there is no alarm)

Message information

Up to 104×1050 (varies depending on the number of messages)

Alarm information

Up to 32×5000 (varies depending on the number of alarms)

N is the number of channels (measurement channels + external input channels + computation
channels).

Example 1: When recording measured data of 30 measurement channels, 240

external input channels, and 10 computation channels with no messages
or alarms.

216 + (88Ч280 + 32) + 3,488 + (104Ч50 + 32) + 832 + (80 + 32 + 280Ч8 + 16 + 2) + (24 + 8)=

36,842 bytes

Sampled Data Size

• Data Sizes of Display Data and Event Data

Channel

Display Data

Event Data

Measurement channel

4 bytes/channel

2 bytes/channel

External input channel

4 bytes/channel

2 bytes/channel

Computation channel

8 bytes/channel

4 bytes/channel

Time data is added to all channels at each sampling.

Time data

8 bytes/sample

• Data Size per Sample
Display Data
(Number of measurement channels×4 bytes) + (number of external input channels×4

bytes) + (number of computation channels×8 bytes) + 8 bytes (time data)

Event Data
(Number of measurement channels×2 bytes) + (number of external input channels×2

bytes) + (number of computation channels×4 bytes) + 8 bytes (time data)