Settings – Yokogawa Button Operated MV2000 User Manual

Page 86

Advertising
background image

3-12

M-4660

Settings

• First-CH, Last-CH

Select the target channels. All of the alarm settings selected here will be applied to

these channels.

• Alarm > 1, 2, 3, 4

Select On for the Alarms that you want to use.

• Alarm > Type

Select the alarm type.

Setting

Name

Description

H

High limit alarm

Σ

L

Low limit alarm

Σ

h

Difference high limit alarm

You can use this alarm on measurement
channels set to difference computation.

l

Difference low limit alarm

You can use this alarm on measurement
channels set to difference computation.

R

High limit on rate-of-change alarm Σ

r

Low limit on rate-of-change alarm

Σ

T

Delay high limit alarm

Σ

t

Delay low limit alarm

Σ

The alarms for channels set to difference computation can be activated by the values

at the positions indicated below.

Input value

Measured value on the reference channel

H, L, T, t

R, r, h, l

Measured value

-

• Alarms > Value

Set the value depending on the type of alarm.

When the Channel Mode Is Volt, TC, RTD, or DI

Type

Value

Alarm Value Range Example

H, L

Within the measurable range.

–2.0000 to 2.0000 V for 2 V range

R, r

Min: The smallest number expressible, given the
decimal place. Max: The value that is equal to the
width of the measurable range (but cannot exceed
30000 after removing the decimal point)

0.0001 to 3.0000 V for 2 V range
0.1 to 1760.0°C for type R
thermocouples

T, t

Same as H and L

Same as H and L

When the Channel Mode Is Delta

Type

Alarm Value

Alarm Value Range Example

H, L

Within the measurable range.

–2.0000 to 2.0000 V for 2 V range

h, l

Within the measurable range.

–1760.0 to 1760.0°C for type R
thermocouples

R, r

Min: The smallest number expressible, given the
decimal place. Max: The value that is equal to the
width of the measurable range (but cannot exceed
30000 after removing the decimal point)

0.0001 to 3.0000 V for 2 V range
0.1 to 1760.0°C for type R
thermocouples

T, t

Same as H and L

Same as H and L

When the Channel Mode Is Scale, Sqrt, or 1-5V

Type

Alarm Value

Alarm Value Range Example

H, L

From –5% to 105% of the scale width.
But within the range of –30000 to
30000, ignoring the decimal point.

–5.0 to 105.0 for a scale of 0.0 to 100.0.
–120.00 to 300.00 for a scale of –100.00 to
300.00.

R, r

Within the range of 1 to 30000,
ignoring the decimal point.

0.1 to 3000.0 for a scale of 0.0 to 100.0.
0.01 to 300.00 for a scale of –100.00 to 300.00.

T, t

Same as H and L

Same as H and L

3.7 Setting Alarms on Channels

Advertising
This manual is related to the following products: