2-14

M-4660

2.4

Connecting the Power Supply

Connecting the Power Cord (if the power supply voltage suffix code is -1)

• Precautions to Be Taken While Connecting the Power Supply

Make sure to follow the warnings below when connecting the power supply. To prevent

electric shock and damage to the MV, observe the following warnings.

WARNING

• Make sure that the power supply voltage matches the MV rated supply voltage

and is within the maximum voltage range specified for the power cord.

• Confirm that the power switch is OFF before connecting the power cord.
• To prevent electric shock, be sure to use a power cord provided by Omega for

use with the MV.

• To prevent electric shock, make sure to ground the MV. Insert the desktop power

cord into a grounded 3-prong outlet.

• Do not use an ungrounded extension cord. If you do, the device will not be

grounded.

Use a power supply that meets the following conditions:

Item

Condition

Rated supply voltage

100 to 240 VAC

Operating supply voltage
range

90 to 132, 180 to 264 VAC

Rated supply frequency

50/60 Hz

Power supply frequency range 50/60 Hz ± 2%
Maximum power consumption

RD-MV1000:45 VA (100 V), 60 VA (240 V)
RD-MV2000:65 VA (100 V), 90 VA (240 V)

Note

Avoid using a power supply voltage of 132 to 180 VAC with the MV because such a power
supply voltage may reduce measurement accuracy.

• Connection Procedure

1.

Check that the power switch is OFF.

2.

Connect the MV power cord to the power inlet on the rear panel.

3.

Make sure that the power outlet meets the conditions listed above and that the

power supply voltage is within the maximum voltage range specified for the power
cord. Then, plug the power cord into the power outlet. Use a grounded 3-prong
outlet.