Yokogawa DAQWORX User Manual

Page 123

Advertising
background image

6-2

IM WX13-01E

Code

Message

Corrective Action

E4626

Cannot shut down Windows while
AddTrigger is running.

Shut down or log off after exiting Launcher.

E4627

A project with the same name already
exists.

A name of a registered project cannot be used. Create a project
with a different name.

E4633

Failed to start the logger.

The program was not correctly installed, or copied files were

E4634

Failed to start the monitor server.

corrupted or deleted during installation. Reinstall. You can uninstall.

E4635

Failed to start the file processor.

the program without losing saved project information.

E4637

Failed to start Data Monitor.

E4638

Failed to start the data viewer.

E4640

Invalid project name.

Project names that start with a period is not allowed. Enter a
correct project name.

E4643

Could not create data file output folder.

Check the settings for the folder specified for data file output.

E4653

No effective logging channels.

Recording cannot be executed if channels are not assigned in the
logging settings or if channels are assigned but valid channels do
not exist. Assign a valid channel.

E4654

Recording was stopped since a monitor
server setting may have been changed.

A setting was changed on the destination data logging. To restart
recording, carry out the start recording procedure again on the
Launcher.

E4656

Failed to open the data file.

Check whether the destination folder exists, write permissions were
granted for the destination folder, sufficient disk space is available,
and whether the file system is operating properly.

E4657

Failed to write to the data file.

Check whether the folder name is valid, write permissions were
granted for the destination folder, sufficient disk space is available,
and whether the file system is operating properly.

E4658

Invalid setting.

The setting file was not created by AddTrigger, or it was created by
a different version of AddTrigger. Install the most recent version of
AddTrigger.

E4659

Monitor server has no data.

Start data logging on the destination data logging software.

E4660

The monitor server is busy.

Start sending data on the monitor server in the data logging
software, then log data using AddTrigger.

E4661

Monitor server stopped data logging.

Data logging on the destination data logging software was stopped,
or data distribution stopped. To restart, start data logging on the
data logging software, or start data distribution. Then, start logging
using AddTrigger.

E4662

No effective recording channels.

Recording/scanning cannot be executed if channels are not
assigned in the logging settings or if channels are assigned but
valid channels do not exist. Check the logging settings, and assign
valid recording channels.

E4663

Failed to reconnect the monitor server.

Couldn’t reconnect during a given time (120 seconds) after the
connection is dropped. Check whether the monitor server of
the data logging software is running and whether the network is
connected properly.

E4664

Port number for Data Monitor is already
in use.

The port number that is displayed when you choose Port No. for
Data Monitor from the File menu on the Launcher is already in use
by another program. Enter an unused port number.

E4665

The new port number for Data Monitor
will be effective after restarting the server.

To make the new port number for Data Monitor effective, restart the
Launcher.

E4668

Failed to convert the data files.

Failed to automatically convert the file. Check whether the
destination folder exists, write permissions were granted for the
destination folder, sufficient disk space is available, and whether
the file system is operating properly.

E4669

Could not create conversion file output
folder.

Check the settings for the folder specified for conversion file output.

E4683

Connection failed.

Check whether the monitor server of the destination data logging
software is running and whether the network is connected properly.