1-10

IM WX13-01E

1.1 Overview of Functions

Terminology for the Period of Data That Is Saved When Setting the
Pretrigger

Data held before the start trigger

The area held in the internal memory that is prior to the point when the start trigger

condition is met.

Pretrigger width
The period including the number of pretrigger data specified in the start conditions.

Pretrigger width = scanning interval x recording rate x pretrigger count.

Data saving period

The period during which data is saved.

Start trigger point

The point in time at which the start condition changes from being unmet to being met.

Stop trigger point

The point in time at which the stop condition changes from being unmet to being met.

Start point of data saving
The first data of the data saving period.
Specified width for data saving
The period including the number of data specified in the stop conditions.
Specified width for data saving = scanning interval x recording rate x data count.

When the Stop Condition Is Set to Alarm Trigger, Level Trigger, or Fixed
Time

When the stop condition is set to Alarm Trigger, Level Trigger, or Fixed Time, data are

saved from the number of data specified for the pretrigger that are prior to the start

trigger point (start point of data saving) up to the start trigger point. Then, scanning is

performed on the data from the start trigger point up to the point specified in the stop

conditions, and that data is saved. In this case, the relationships between the data held

before the start trigger and the pretrigger width are shown in the figure below.

Data held before the start trigger > pretrigger width (most common

case)

Data held before the start trigger

Pretrigger width

Start point of data saving

Start trigger point

Stop trigger point

Time

Data saving period

Data held before the start trigger < pretrigger width

Data held before the start trigger

Pretrigger width

Start point of data saving

Start trigger point

Stop trigger point

Time

Data saving period