Quarterly maintenance annual maintenance coils, Service – Mammoth 800 to 5500 CFM: Dedicated Outdoor Air System with Enthalpy Wheels (VHC) User Manual

Page 15

Advertising
background image

MAMM-VHC-IOM-1A (JANUARY 2013)

15

Quarterly maintenance (every three months) should include:

Air Filters

The standard medium efficiency filters and optional high ef-

ficiency filters are disposable and should be replaced every

three months. More frequent replacement may be required

under extremely dirty operating conditions.
To replace the filters, open the filter access door, grasp the

filters and pull straight out. The filters will slide completely

out of the unit. Slide the new filters into the frame and close

the filter access door.

Cassette Panels and Interior of Unit

Remove the filters from the unit. Wipe the foil faced insu-

lation, or the optional interior galvanized liner, surfaces

and cassette panels with a soft cloth and mild cleaning

solution.

Annual maintenance should include:

Aluminum Enthalpy Wheel

No cleaning of the enthalpy wheel is required as it is

self-cleaning due to the opposing airflows. If it is desired

to clean the enthalpy wheel, use low pressure air or a

vacuum. Wash the cassette panels with a soft cloth and

mild cleaning solution. Visually inspect the cassette brush

seals (shown in Figure 4), perimeter seal and drive belt

for proper operation.

Fans

Blower wheels and fan housing should be checked for dirt

buildup. If they are dirty, it will be necessary to remove the

blower assembly to clean the dust out through the fan mouth.

System Operation Check

Verification of all control modes should be checked to en-

sure proper operation. Refer to

Start-up

section.

Dirt on the surface of the coil reduces its ability to transfer

heat which lowers the efficiency of the unit, resulting in

poor air quality and expensive operating costs. Because

of the condensate on the coil, the dirt often becomes wet

and contributes to the growth of microbial organisms.

Negligence in maintenance may result in serious health

related indoor air quality problems.
The coil should be kept clean for maximum performance.

To achieve maximum efficiency, clean the coil often during

periods of high demand or when dirty conditions prevail.

Mammoth recommends cleaning the coil a minimum of

once per year to prevent dirt buildup in the coil fins where

it may not be visible.
Coil fins can be cleaned by using steam with detergent, hot

water spray or a commercial chemical coil cleaner. After

cleaning the coil, be sure to rinse thoroughly.

Cleaning Procedure

1. Shut down the unit by closing the main disconnect

at the power inlet.

2. Open panels or doors to gain access to both sides of

the coil section.

3. Remove soft debris from both sides of the coil with a

soft brush.

Quarterly Maintenance

Annual Maintenance

Coils

Rotor

Brush seal

Figure 4: System operation check

Service

CAUTION

Do not use acidic chemical coil cleaners. Do not use al-

kaline chemical coil cleaners with a pH value greater than

8.5 or lower than 6 (after mixing) without using an alu-

minum corrosion inhibitor in the cleaning solution. Using

these types of cleaners may result in unit damage.

Advertising