CyberScan pH5500/ 5000

Direct Reading with Blank Offset

This method presents you with the ability to measure samples with
low level concentrations of the ion of interest. It eliminates
background interference by subtracting the value of a blank
concentration from the value for the sample. See ISE theory pg 133.

To Standardize the Meter

You will need to standardize the meter with one blank and a minimum
of two standards.

1. Touch

STD on the Ion measure screen to access the standardize

screen.


2. Touch

clear to delete the previous standards.


3. Add ionic strength adjuster to the standards and the blank.

4. Immerse the ion selective electrode in the standard containing

the lowest concentration of the ion of interest.


5. Stir the standard and/ or blank.

6. Touch

std again to initiate the standardization process. At this

point, the meter will display the question” is this standard or a
blank?”


7. Touch

STND if the sample that the electrode is immersed in is an

Ion standard solution.


OR

Touch

BLANK if the sample that the electrode is immersed in is

the blank solution.


Remember, HELP is always just a touch of the button away.