Thermo Fisher Scientific CyberScan pH 5000 User Manual

Page 51

Advertising
background image

Instruction Manual

CyberScan pH5500/ 5000

This setup option allows you to select from 3 different buffer groups
each containing 5 buffers, for auto buffer recognition. Or you can
create a custom group of buffers for auto buffer recognition by
touching custom.


The 3 existing buffer groups are:

USA buffers:

2,4,7,10 and 12

European buffers:

1,3,6, 8 and 10

NIST buffers:

1.68, 4.01. 6.86. 9.18 and 12.45

To Select Buffer Group

1. Access the select pH buffer group screen from the pH Setup

screen. The current buffer group is displayed on the screen.

2. Touch USA or NIST or EURO or custom on the right of the

screen to select a buffer group

3. Touch enter to accept the buffer group to be used for auto

recognition.

OR

Touch exit to return to pH Setup Screen, without making any
changes.

46

Advertising
This manual is related to the following products: