Instruction Manual
CyberScan pH5500/ 5000
Sequential ID# Assignment
1. Access the Set Sample ID# screen from the pH (mV, Ion) Setup
screen.
2. Touch seq for sequential ID# assignment. The current ID# is
displayed on the screen.
3. Touch
clear to delete the current ID#.
4. Use the alphanumeric keypad on the screen to enter the number
that you would like your sequential ID# assignment to begin with.
Every time you touch print on the measure screen, the ID# will
increase by 1. The BS key will allow you to backspace to remove
a character that was incorrectly entered.
5. Touch
enter to accept the first sequential ID# and return to the
pH (mV, Ion) Setup screen.
OR
To Deactivate the Sample ID# Assignment
1. Access the Set Sample ID# screen from the pH (mV, Ion) Setup
screen.
2. Touch
man for manual ID# entry. The current ID# is displayed on
the screen.
3. Touch
clear to delete the current ID#.
4. Touch
enter. The ID# assignment is now deactivated. No number
will be assigned to your samples. The meter will return to the pH
(mV, Ion) Setup screen.
OR
Touch exit to return to the pH (mV, Ion) Setup screen, without
making any changes.
44