Instruction Manual

CyberScan pH5500/ 5000

To Standardize the meter with Auto Buffer Recognition

1. Touch STD on the pH measure screen to access the

standardized screen


2. Touch clear to delete a previous standardization. All pH

standardization values will be cleared. The meter will remain on
the standardization screen at this point.


OR

If the screen says “Not standardized” proceed to step 3.


3. Immerse your rinsed electrode (s) in a buffer solution from the

selected buffer group that you chose during the setup process
(see page 47) and stir gently.


4. Touch

std again to standardize the meter using this buffer. The

word MEASURING will flash until the signal is stable. The meter
will accept the value once it recognizes that the reading is stable.
At that time, STABLE will appear on the screen at this time. A
beaker icon and the entered pH value will also appear on the
screen at this time. The meter will then return to the measure
screen.


5. Touch std on the pH measure screen to access the

standardization screen and repeat steps 3 through 5 to
standardize with up to 5 buffers.


Updating a standardization with Auto Buffer Recognition

You can update a standardization at anytime. To update a
standardization, place your electrode in any of the buffers originally
used to standardize the meter and touch std on the touch screen
twice. The value for that buffer will be updated and stored in the
meter’s memory. The time of standardization will also be updated on
the pH measurement screen.

The efficiency of the electrode is reported as the slope. When doing a
multi-point standardization, the slopes of the individual segment are
calculated by the meter. The slope that appears on the screen is the
slope that is the least perfect, or the farthest from 100%.

For optimal results, the meter should be standardized at a minimum
of every 8 hours. For more accurate measurements, the meter should
be standardized more frequently.