13 TROUBLE
SHOOTING
Your CyberScan meter contains many error messages to provide aid
should trouble occur with a measurement or meter operation (touch
pad and input errors). The messages are accompanied by a brief
description of the error, and in some cases advice on how to correct
it. An example of an error message is:
Bad Electrode Slope
The electrode you have standardized has a slope which is out of the
normally acceptable range of 90% to 102%. You should try to
restandardise, or replace your electrode if the problem persists. Note
that you can continue to make measurements with this electrode;
however, the readings may be inaccurate.
Other error messages delivered by the meter are:
Data Log full
Data log empty
Unrecognized
date
Unrecognized
time
Invalid pH limit
Invalid ion limit
Invalid
limit
Limit
exceeded
Channel 1 limit exceeded
Channel 2 limit exceeded
Invalid
temperature
Invalid
Isopotential
Invalid Print interval
Invalid pH value
Sample ID not found
Operator ID not found
Also, whenever possible, touch help for complete information about