13 TROUBLE

SHOOTING

Your CyberScan meter contains many error messages to provide aid
should trouble occur with a measurement or meter operation (touch
pad and input errors). The messages are accompanied by a brief
description of the error, and in some cases advice on how to correct
it. An example of an error message is:

Bad Electrode Slope

The electrode you have standardized has a slope which is out of the
normally acceptable range of 90% to 102%. You should try to
restandardise, or replace your electrode if the problem persists. Note
that you can continue to make measurements with this electrode;
however, the readings may be inaccurate.

Other error messages delivered by the meter are:

Data Log full

Data log empty

Unrecognized

date

Unrecognized

time

Invalid pH limit

Invalid ion limit

Invalid

limit

Limit

exceeded

Channel 1 limit exceeded

Channel 2 limit exceeded

Invalid

temperature

Invalid

Isopotential

Invalid Print interval

Invalid pH value

Sample ID not found

Operator ID not found


Also, whenever possible, touch help for complete information about
the meter operation in which you are currently engaged,