6.6 Select Auto Read Mode
You can use this meter when the Auto Read Mode function is active
or when it is inactive. When the Auto Read function is active, the
meter will lock onto a reading when the meter recognizes it as stable.
The meter will not deviate from this reading until meas is touched. If
the auto Read function is inactive, then the meter will continuously
monitor the pH of the sample and the Measure screen display will
indicate any fluctuation in the sample pH.
To Select Auto Read Mode
1. Access the Select Auto Read Mode screen from the pH Setup
screen. The current Read Mode id displayed on the screen.
2. Touch
AUTO or MAN to choose the desired read mode.
3. Touch
enter to accept the read mode and return to the pH setup
screen.
OR
Touch exit to return to the pH Setup Screen, without making any
changes.
enter
Current MODE
Set pH Auto Read Mode
AUTO
MAN