15.2 Data Management



Your CyberScan meter is equipped with an RS-232 output port for
sending measurement readings to a printer or computer. To use the
meter with a printer or a computer, the printer or computer must have
an RS-232 part, also. It should be capable of receiving data within the
following guidelines.

Baud rates

110 to 38400 BPS

Data bits

7 or 8

Parity

even, odd or none

Stop bits

1 or 2


The meter and the output device must be configured to match with
respect to these parameters.

A shielded RS 232 cable must be used. The RS 232 connector is on
the back of the meter. (See diagram page 8) the connector is a
female DB9 style connector. If the output device has a 25 pin
connector, an adapter is required. If the meter and the printer or
computer are to work together, the pin configuration must match. The
meter’s 9 pin configuration is as follows.

Pin

Description

1

Buffers analog copy of pH input 1

2

Send data from meter

3

Receive data to meter

5

Signal

Common

9

Signal

Common


To initiate printing, touch print if the manual print mod, or set the print
interval for stable readings or at a prescribed timed interval. Refer to
the setup sections. Set print interval and set print criteria.

To download data stored in memory to a printer or computer, access
the Viewed Stored Data option in the setup screens. Access the
desired data point by sorting the data by one of the three options.
Touch print on the screen displaying the data.