CyberScan pH5500/ 5000

To take a measurement

1. Immerse the ion selective electrode in a known volume of the

sample solution.


2. Stir the sample.

3. Touch

KA to take the first mV reading of sample. Wait for the

reading to stabilize (STABLE will appear).


OR

Touch

KA to accept the current reading.


4. Add a known volume of standard solution to the sample.

5. Touch

KA to take the second mV reading of sample. Wait for the

reading to stabilize (STABLE will appear).


OR

Touch

KA to accept the current reading.


6. Use the keypad to enter the sample volume.

7. Touch enter to accept the value.

8. Use the keypad to enter the standard volume.

9. Touch

enter to accept the value.


10. Use the keypad to enter the concentration of the standard

solution added to the sample.


11. Touch enter to accept the value.

12. Use the keypad to enter the concentration of the standard

solution added to the sample


13. Touch enter to accept the value. Your Known Addition result is

now displayed.



14. Touch print to send the data to memory if sample ID# has been

assigned or to print the data on the screen.


OR

Touch

OK to return to the Measure screen for a new

measurement.