Thermo Fisher Scientific CyberScan pH 5000 User Manual

Page 120

Advertising
background image

Instruction Manual

CyberScan pH5500/ 5000

11.4 Known Subtraction Method


The procedure for known subtraction is very similar to that for known
addition method. An aliquot of standard is added to a known volume
of the sample. The difference is that the standard does not contain
the same ionic species that you are trying to measure in the sample.
Instead, it contains an ion that will complex or precipitate the ion of
interest, removing it from the sample. Like known addition, the mV
readings are taken before and after the standard is added to the
sample and the difference in mV values can be used to calculate the
concentration of free ions of interest in the sample. This method will
also require you to enter values of certain parameters in order to
obtain an ion concentration. See ISE theory page 133.

To take a measurement

1. Immerse the ion selective electrode in a known volume of the

sample solution.


2. Stir the sample.

3. Touch

KS to take the first mV reading of sample. Wait for the

reading to stabilize (STABLE will appear).


OR

Touch

KS to accept the current reading.


4. Add a known volume of standard solution to the sample.

5. Touch

KS to take the second mV reading of sample. Wait for the

reading to stabilize (STABLE will appear).


OR

Touch

KS to accept the current reading.


6. Use the keypad to enter the sample volume.

7. Touch enter to accept the value.

8. Use the keypad to enter the standard volume.

9. Touch

enter to accept the value.

115

Advertising
This manual is related to the following products: