Option: Connect the optional FET electrode by plugging it into
the FET jack on the back meter panel. Allow the FET electrode to
warm up five minutes before use.

Connect ion specific electrodes in the same manner as pH
electrodes.

If both a conventional and a FET electrode are connected to the
meter, do not put them in a solution together because you will get
inaccurate measurements.
Do not discard the BNC shorting cap.


3. Rinse and blot-dry (do not wipe)
electrodes between each
measurement.

Rinse

electrodes

with distilled or deionized water,

or a portion of the next solution

to be measured.








4. Between

measurements,

store

conventional

pH

electrodes

storage solution, pH 4 buffer, or

KCI solution. Always leave the

filling hole of liquid filled

combination electrodes open.

Refill

when the level of

solution gets below the

manufacturer’s

recommended

level.


Proper electrode care is fundamental to obtaining reliable pH
measurements. Improper care of the electrode may cause the meter
reading to drift, respond slowly, or produce erroneous readings. For
this reason, the electrode should always be conditional and used in
accordance with manufacturer’s instructions.