Maintenance plan, 4 maintenance plan – INFICON UL1000 Helium leak detector User Manual

Page 101

Advertising
background image

Ma

intena

nce Work

101

iina70e 09.fm

technical handbook

(1408)

9.4

Maintenance Plan

Subassembly

Maintenance work

Operating hours/years

Service

level

Spare part

UL1000 / UL1000 Fab

1500

4000

8000

16000

24000

number

1/4

1

2

3

4

Vacuum system

Backing pump D16 B

Check oil level, if required exchange

X

I & II

Exchange the oil

2

X

II

20099091

Refurbish backing pump

X

III

Scroll pump Agilent

Exchange the tip seal

X

1

III

200001671

TS 620

Exchange the scroll head

X

III

200001665R

SplitFlow 80

Exchange the operating agent reservoir

X

2

II u. III

200003801

Bearing exchange

X

2

III

Valve block

Clean the valves, replace valve seals

2

X

III

200000594

Disassemble and clean the valve block

2

X

III

200000593

Replace filter of vent and purge gas line

1

X

1

I, II ,III

200000683

Align Pirani gauge

X

III

Silencer UL1000 Fab

Replace

X

1

I, II, III

20099183

Exhaust filter UL1000

Check, empty

X

I, II, III

Replace filter insert

X

1

I, II, III

200000694

Electrical system

Fan subassemblies

Blow out fan chassis wall and bottom

1

X

1

I

Replace replacement filter cell for chassis
wall fan

1

X

1

I

200000685

Advertising
This manual is related to the following products: