Accessories 28, Digital out, 1 accessories – INFICON UL1000 Helium leak detector User Manual

Page 28: 2 digital out

Advertising
background image

28

Installation

iina

70

e 0

2

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

2.3.2.1

Accessories

The following accessories may be connected to the sniffer line SL200 (Please refer
to Chapter

Fig. 2-6

/1) or the test chamber TC1000:

Contact pins 1 and 3 are fused with a 0.8 A slow-blow fuse. The amount of power
which can be drawn is limited to 10 W. The contacts are numbered from top to
bottom.

2.3.2.2

Digital Out

The following relay outputs are available for further signal processing. The maximum
rating for the relay contacts is 60V AC/1A.

Description of the operation mode of the Digital Out.

Trigger 1:
Is open in case Trigger Level 1 is exceeded or the machine is not in condition of
measuring.

Trigger 2:
Is open in case Trigger Level 2 is exceeded or the machine is not in condition of
measuring.

Zero active:
Is closed in case Zero function is running.

Ready:
Is closed in case machine is ready for measurement (Emission on, no error).

CAL active
Closed when machine is in calibrating routine.

Pin

Assignment

1

+24 V, constantly applied, power supply for the sniffer line SL200.

2

GND24

3, 6

Input

4, 5, 7, 8

Output

Pin

Assignment

1

+24V, bridged with pin 1 of socket „Digital In“

2

GND_24V

3

Trigger 1

4

Trigger 2

5

Free

6

Zero active

7

Ready

8

CAL active

9

Cal request

10

Error

11

Warning

12

Purge

13

Measure

14

Recorder Strobe

15

Common dig. out

16

Free

Advertising