Fig. 6-14 – INFICON UL1000 Helium leak detector User Manual

Page 70

Advertising
background image

70

Description of the Menu

iina

70

e 0

6

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

The measurement period (Maximum value) depends on the volume of the chamber,
volume of the tested object and the rejectant leak rate.

If the measured leak rate has gone below the trigger level, the measurement period
will be finished with the result PASS even if the maximum value of the measurement
period is not reached.

Examples for time settings
(using the Inficon Test Chamber with a volume of 430 cm³):

*external calibration with a 10E-9 test leak ( i.e. TL 9 ) recommended

Notice: After changing the measurement time a calibration request could be asked.

Settings:

Interval:

time

1 - 20 sec

1 sec steps

20 - 30 sec

2 sec steps

30 - 60 sec

5 sec steps

1 - 30 min

10 sec steps

3 - 10 min

30 sec steps

10 - 30 min

1 min steps

Fig. 6-14 Measurement period

Range of rejectant level

Measurement Period

10E-5

2 sec

10E-6

2 sec

10E-7

2 sec

10E-8

>5 sec

10E-9

>10 sec*