Save parameter set, Monitoring functions, Calibration request particle protection – INFICON UL1000 Helium leak detector User Manual

Page 82: 7 monitoring functions

Advertising
background image

82

Description of the Menu

iina

70

e 0

6

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

6.6.6.2

Save parameter set

Main Menu

>

Settings

>

Parameter save / load

>

Load parameter set

The settings of the selected saved parameter set will be displayed and can be
reloaded.

6.6.7

Monitoring functions

Main Menu

>

Settings

>

Monitoring functions

Calibration request

Main Menu

>

Settings

>

Monitoring functions

> Calibration request

It can be selected whether the operator is reminded of the fact that a calibration may
has become necessary or not. The default value is off.

If the calibration request is switched on, a corresponding message will appear when
30 minutes have elapsed after power on or if the temperature of the UL1000 and
UL1000 Fab has changed by more than 5 °C (9 °F) since the last calibration.

Particle protection

Main Menu

>

Settings

>

Monitoring functions

> Particle Protection

This mode can be switched on and off.

If switched on the UL1000 and UL1000 Fab will not start pumping down before the
inlet ha

s not

dropped below

1

mbar. e.g. it is assumed that the part under test is

pumped by another pump in parallel.

Purpose: When the leak detector does not pump at high pressures no gas stream

,

possibly carrying particles gets into the leak detector.

Softkey 6:

Arrow up

Upward to the previous screen.

Softkey 7:

Arrow down

Downward to the next screen.

Softkey 3:

Off

The calibration request will be switched off.

Softkey 7:

ON

The calibration request will be switched on.

Softkey 3:

Off

Softkey 5: Help

Softkey 7:

ON

Advertising