Units

Main Menu

>

Trigger & Alarms

>

Units

The prefered leak rate unit can be selected. There is the choice of 4 (mbar, Pa, Torr,
atm)pressure units and 5 leak rate units (mbar l/s, Pa m

3

/s-1, Torr l/s, atm cc/s).

Notice: In Sniff mode the following measuring units are electable (Refer to chapter

6.3

): ppm, g/a eq (helium leak rate is equivalent with leak rate R134a), oz/

gr eq (helium leak rate is equivalent with leak rate R134a).

6.4.5

Alarm delay

Main Menu

>

Trigger & Alarms

>

Alarm delay

In some applications (for instance during pump down in a „chamber test system“) it
might be necessary to block an alarm for some time after pressing START.

This delay time of the alarm can be changed.

After pressing START the loudspeaker is activated as soon as the leak rate drops
below trigger level 1 or after the entered alarm delay time has elapsed. This setting
is only active for the audio alarm types SETPOINT and TRIGGER ALARM (See
Chapter

6.4.6

).

Softkey 2:

Arrow up

Scroll up to select a pressure unit.

Softkey 3:

Arrow down

Scroll down to select a pressure unit.

Softkey 6:

Arrow up

Scroll up to select a leak rate unit.

Softkey 7:

Arrow down

Scroll down to select a leak rate unit.

Softkey 3:

Arrow down

Decrease the delay time. The minimum value is 0 seconds.

Softkey 7:

Arrow up

Increase the delay time. The maximum value is 10 minutes up to infinity.