Lc display, Start button, Stop button – INFICON UL1000 Helium leak detector User Manual

Page 41: Zero button, S the, Ess the

Advertising
background image

Description and Working Principle

41

iina

70

e 0

4

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

4.2.2.1

LC Display

The

LC Display

Fig. 4-2

/1 is the communication interface to the operator. It displays

the leak rates, the status report of the machine, messages, warnings and errors.

4.2.2.2

START Button

Pushing the

START Button

Fig. 4-2

/6 enables the UL1000 and UL1000 Fab to start

the measure procedure. If the START button is pushed again in measurement mode,
the maximum leak rate indicator („hold“ function) is acitvated. This indicator shows
the maximum leak rate since „START“. By pressing the START-button again the
„hold“ function will be started.

4.2.2.3

STOP Button

Pushing the

STOP Button

Fig. 4-2

/13 interupts the measure procedure. If the button

is pressed longer the inlet is vented according to the conditions defined in the menu

Vent delay

. Please refer to chapter

6.6.1.2

Vent delay

to select the time parameters

of the venting.

4.2.2.4

ZERO Button

Pushing the

ZERO Button

Fig. 4-2

/14 enables the zero mode.

When pressing ZERO the currently measured leak rate is taken as a background
signal and is subtracted from all further measurements. As a result the displayed leak
rate then is

To reverse the ZERO function please keep the button pressed for about 3 seconds.

After pressing ZERO the decreasing background is fitted to the course automatically.
So it is possible to recognize leaks even when the signal is decreasing rapidly.

Please also refer to the pictures below.

1×10

-6

in GROSS

1×10

-10

in FINE

1×10

-12

in ULTRA