Trigger & Alarms

Main Menu

>

Trigger & Alarms

The trigger levels, the volume of the loudspeaker and the units of leak rates and
pressures can be set in this menu

Fig. 6-9

.

6.4.1

Trigger Level 1

Main Menu

>

Trigger & Alarms

>

Trigger Level 1

The value of the first trigger level can be set. See Chapter

4.2.2.7

Numerical Entries

for the description of the entry.

Trigger 1 and Trigger 2 are programmable switching thresholds. When these
thresholds will be exceeded the UL1000 and UL1000 Fab reacts as follows:

Display

In the status line of the display the signs for Trigger 1 and Trigger 2 are displayed
inverted if the leak rate exceeds (becomes higher than) the programmed value.

Relay Output

The trigger-relais of the digital out switches. Please refer to Chapter

2.3.2.2

, Digital

out, for further details.

Alarm/Loudspeaker

Additionally Trigger level 1 defines at which level the various alarm types react (See
Chapter

6.4.6

,

Audio alarm type

)

Fig. 6-9: The

Trigger & Alarms

Menu

Key No.

Name

Description

1

Back

Return to the main menu.

2

Trigger Level 1

Definition of Trigger level 1. See chapter

6.4.1

3

Trigger Level 2

Definition of Trigger level 2. See chapter

6.4.2

4

Volume

See chapter

6.4.3

5

Units

Selection of leak rate and pressure units.
See chapter

6.4.4

6

Not used in this menu.

7

Alarm delay

See chapter

6.4.5

8

Audio alarm type

Choice of different alarm types. See chapter

6.4.6