List of Errors & Warnings

The following pages contain a list of all errors and warnings displayed at the control
panel. Warning messages are indicated by numbers with a leading W. Error
messages are indicated by numbers with a leading E

.

No.

Displayed Message

Description and possible solutions

E04

Temperature monitoring,
turbo molecular pump is
defectt

Short circuit in the temperature sensor

E05

Temperature monitoring,
turbo molecular pump is
defect

Temperature sensor disconnected

W15 Leakrae is too high! Machine

switched into stand-by to
prevent contamination.

The survey function „contamination“ is activated. A leak rate higher than
the adjusted value was detected.

Gross leak

Switch off limit is set too low

Alarm delay time too short

W16 Turbo molecular pump

service interval expired!

The service intervall for the turbo molecular pump is expired.

W17 Forepump service interval

expired!

The service intervall for the fore pump is expired.

W18 Exhaust filter service interval

expired!

The service intervall for the exhaust filter is expired.

W21 EEPROM write time out

EEPROM defective
MC 68 defective

W22 EEPROM parameter queue

overflow

EEPROM defective
MC 68 defective

E23

24V of the OPTION socket is
too high

The tension 24V at socket OPTION is too high.

E24

24V at socket OPTION is too
low.

Fuse F2 on the I/O board has blown

E25

Receded valve voltage too
low (< 7V).

I/O board is faulty.

W28 Real time clock reset! Please

enter date and time!

Battery at MC68 is discharged or faulty.

MC68 had been replaced.

E29

24V supply for fans ist too low
(< 20V).

Fuse F1 on wiring backplane has blown.

E30

24 V of the remote control is
too low (> 20V).

Fuse F1 on the I/O-board has blown.

W31 The offset voltage of the

preamplifier is too high
(> 5mV).

The preamplifier is faulty.

W32 Preamplifier temperatur is too

high (> 60°C).

Ambient temperature is too high.

Air filter dirty.

W33 Preamplifier temperature is

too low (< 2°C).

Ambient temperature is too low.

Temperature sensor is faulty.