Default parameters, 5 default parameters – INFICON UL1000 Helium leak detector User Manual

Page 33

Advertising
background image

Installation

33

iina

70

e 0

2

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

2.5

Default parameters

The following parameters are set like shown when in the menu of the UL1000 and
UL1000 Fab under Settings

 Parameters load/save, „load default values“ is

chosen.

Auto-scaling:

On

Scaling

logarithmic

Display range:

4 decades

Time axis:

32 seconds

LCD invers:

OFF

Background in stand by mode:

OFF

Automatic calibration request:

OFF

Mass:

4 (helium)

Recorder Output:

leak rate

Volume:

2

Leak rate unit:

mbar l/s

Mode:

Vacuum

Trigger level 1:

1E-9 mbar l/s

Trigger level 2:

1E-8 mbar l/s

Leak rate external test leak (Vacuum):

1E-7 mbar l/s

Leak rate external test leak (Sniffer):

1E-5 mbar l/s

Vent delay:

2 seconds

Automatic purge:

OFF

Pressure unit:

mbar

Minimum volume:

0

Beep:

ON

Maximum evacuation time:

30 minutes

Audio Alarm Typ:

Trigger Alarm

Maximum inlet pressure when sniffing:

1 mbar

Minimum Inlet pressure when sniffing:

0,1 mbar

Number of decimal place at leak rate displayed:

1

Scroll display:

On

Particle protection:

Off

Direct access to calibration:

On

Contamination protection:

Off

Switch off limit for contamination protection:

1E-3 mbar l/s

Control location:

Local

Alarm delay:

30 seconds

Leak rate filter:

I•Cal

Zero:

enabled

Advertising