6.4.6

Audio alarm type

Main Menu

>

Trigger & Alarms

>

Audio alarm type

The trigger of the audio alarm can be switched on or off.

6.4.6.1

Pinpoint

The tone of the acustical signal changes its frequency only in a LR-window

Fig. 6-10

which ranges from one decade below the Trigger level 1 up to one decade above the
Trigger level 1. Below the window the tone is constantly

low

, above the window it is

constantly high.

Example: The Trigger level 1 is 4×10

-7

mbar l/s. So the window where the tone

changes reaches from 4×10

-8

mbar l/s up to 4×10

-6

mbar l/s.

6.4.6.2

Leak rate prop.

The frequency of the accoustic output is proportional to the reading on the bargraph
display. The frequency ranges from 300 Hz to 330 Hz. Please

refer to Chapter

6.2.1

Scale linear/logarithmic

for the definition of the number of decades.

Softkey 2:

Pinpoint

Use this function to localize a leak with a well-known size. Please refer to Chapter

6.4.6.1

Softkey 3:

Leak rate prop.

The sound will be proportional to the leak rate signal.

Please refer to Chapter

6.4.6.2

Softkey 5:

?

Help

Softkey 6:

Setpoint

Please refer to Chapter

6.4.6.3

Softkey 7:

Trigger alarm

An alarm sounds when the trigger 1 is exceeded.

Please refer to Chapter

6.4.6.4

Fig. 6-10: Pinpoint

f

LR

Tr. 1

0.1 • Tr. 1

10 • TR. 1

2570 Hz -

491 Hz -