6.6.4

Interfaces

Main Menu

>

Settings

>

Interfaces

The parameters of the interface can be set.

6.6.4.1

Control Location

Main Menu

>

Settings

>

Interfaces

>

Control Location

Softkey 3:

Control Location

Please refer to Chapter

6.6.4.1

Softkey 4:

RS232 Protocol

Please refer to Chapter

6.6.4.2

Softkey 7:

Recorder output

Please refer to Chapter

6.6.4.3

Softkey 8:

Scaling Recorder Output

Please refer to Chapter

6.6.4.4

Softkey 2:

PLC

The UL1000 and UL1000 Fab is controlled via the Digital In connector (See Chapter

2.3.2.3

). The START, STOP and ZERO buttons at the control panel are locked.

Softkey 3:

RS232

The UL1000 and UL1000 Fab is controlled via RS232 interface by an external
computer. In this mode the UL1000 and UL1000 Fab can not be controlled via
keyboard. The START, STOP and ZERO button at the machine are deactivated.

Taste Nr. 4

All

UL1000 and UL1000 Fab can be operated via the digital input, the RS232-interface
and the keyboard.

Softkey 5:

Local & PLC

The UL1000 and UL1000 Fab is controlled via the Digital In connector or the
START, STOP and ZERO buttons at the control panel.

Softkey 6:

Local & RS232

The UL1000 and UL1000 Fab is controlled via the Digital In connector or the
START, STOP and ZERO buttons at the control panel.

Softkey 7:

Local

The UL1000 and UL1000 Fab is controlled via the START, STOP and ZERO buttons
at the control panel. The Digital In connector is not used.