Internal calibration, Automatic internal calibration, Manual internal calibration – INFICON UL1000 Helium leak detector User Manual

Page 89: External calibration, 1 internal calibration, 2 external calibration

Advertising
background image

Calibration

89

iina

70

e 0

7

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

7.2.1

Internal Calibration

For internal calibrations the UL1000 and UL1000 Fab differentiates between two
possibilities:

If the unit is blanked off or disconnected from any chamber by a valve on the

Inlet

Port

the automatic calibration can be chosen (Soft Key no. 8).

If the unit is connected to a chamber or a bigger component the calibration has to
be performed manually because the reaction times on opening or closing the
internal leak standard vary depending on the volume of the chamber.

Notice: It is recommended to use the automatic calibration if possible.

7.2.1.1

Automatic Internal Calibration

Once this procedure is started the entire procedure is performed automatically. At the
end (after about 25 s) a beep is released. Thereafter the unit is ready for further use.

7.2.1.2

Manual Internal Calibration

When

Manual Internal Calibration

is selected it is assumed that the UL1000 and

UL1000 Fab is connected to a component under test (if not please go to

Automatic

Internal Calibration

).

After starting the

Manual Internal Calibration

the UL1000 and UL1000 Fab pumps

down the test part (if not already under vacuum) and opens the internal leak
standard. Depending on the volume of the part it may take some time for the helium
signal to stabilize. Therefore the user has to confirm that the signal has reached a
stable level (Soft Key no. 8).

The unit now runs through the tuning process and closes the internal leak standard
automatically. Again the volume of the test part determines how long it takes to pump
down the helium and to reach a stable background level, which has to be confirmed
by the user.

Thereafter the unit is calibrated.

7.2.2

External Calibration

For an external calibration a leak standard has to be attached to the part under test
or the inlet port directly.

Notice: The shown leak rate can diverge of the printed values of the external

calibrated leak because of uncertainities and temperature coefficients.

After

External Calibration

(Soft Key no. 8) has been chosen the following messages

are displayed and the described actions are required: