SETPOINT THICKNESS

Figure 2.10: Set Point Thickness Parameter.

LEGEND

SP THK

RANGE

0 to 9999

UNITS

KILO ANGSTROMS

The SETPOINT THICKNESS parameter is used to provide a comparison

point for the STM-100 / MF Thickness Setpoint Relay. This set of relay
contacts will be closed whenever the thickness display equals or exceeds the
setpoint value. A setpoint value of 0.000 kÅ causes the thickness setpoint
function to be ignored.

SETPOINT TIMER

Figure 2.11: Set Point Timer Parameter.

LEGEND

SP TMR

RANGE

00:00 to 99:59

UNITS

MIN:SEC

The SETPOINT TIMER parameter is used to provide a comparison point

for the STM-100 / MF Timer Setpoint Relay. The internal STM-100 / MF timer
is pre-set with the setpoint value whenever the front panel ZERO key, or a
remote ZERO TIMER input, is activated. The Timer Relay contacts are
opened at this time and the timer display begins counting downward toward
zero. When the timer display reaches zero the relay contacts are closed and
the display timer begins counting in an upwards fashion. A setpoint value of
00:00 will cause the timer setpoint function to be ignored.

Setpoint Thickness
Parameter

Setpoint Timer

Page 2 - 7

OPERATION AND PROGRAMMING