_________________________________________________________________________

Command:

P

Parameter:

None

Description:

Returns the status of the end thickness status. This is not tied to a relay, but to an
annunciator. End Thickness closes the shutter and sets this status. A Zero
Thickness command clears this status.

_________________________________________________________________________

Command:

Q

Parameter:

None

Description:

Returns the status of the 4 remote inputs. A code of @..O is sent back. The lower 4
bits of the character represent the 4 inputs. A 1 indicates that the input is at ground.
A 0 indicates that the input is open or at a logic high. The following table defines the
response bit weights with each input.

Input Name

Active "ON" Weight

Zero Timer

1

Zero Thickness

2

Shutter Close

4

Shutter Open

8

_________________________________________________________________________

Command:

R

Parameter:

None

Description:

Returns the status of the configuration switches. Because there are 12 switches, the
returned value is between 0 and 4095. A value of 0 indicates that all switches are in
the OFF (or down) position. Switch number 12 is theLS (1) bit, switch number 1 is
the MS (2048) bit.

_________________________________________________________________________

Command:

S

Parameter:

None

Description:

Returns the thickness value, in Angstroms. This number will always contain 7 digits
with a leading space or minus sign. Leading zeros are not suppressed.

_________________________________________________________________________

Command:

T

Parameter:

None

Description:

Returns the rate value, in A/s. The format is a leading space or minus followed by
NNN.N.

_________________________________________________________________________

Command:

U

Parameter:

None

Description:

Returns the sensor frequency, in Hz. If the crystal has failed, the number is the last
valid reading, or blanks if there was never a good reading.

Get the Status of the
End Thickness Relay

Get the Remote Input
Status

Get the Configuration
Switch Status

Get the Thickness
Value

Get the Rate Value

Get the Sensor
Frequency