Communication demo disk, Setting baud rates, Table 5.1: baud rate configuration table – INFICON STM-100/MF Thin Film Deposition Monitor User Manual

Page 57

Advertising
background image

STM - 100 / MF

9 PIN MALE

SHIELD

IBM - PC,XT

25 PIN FEMALE

2 RxD

3 TxD

9
5 GND

TxD 2

RxD 3

CTS 5

DSR 6

GND 7

DTR 20

Figure 5.2: Connections for STM-100 / MF to PC Compatible Computers.

After the proper cable is made, make sure the baud rate of the computer and

STM-100 / MF are the same. By setting two of the dip switches on the
CONFIGURATION set of switches, you can set one of four baud rates as shown in
Table 5.1.

The diskette included with this manual contains demonstration software that

will verify the communications connection between a MSDOS based computer and
the STM-100 / MF. The program STMCKOUT.EXE communicates with the STM
via RS232 and will exercise the communication sub system. For more information
of other programs on the diskette refer to Appendix A. Follow the directions in the
next paragraphs to set up the STM for serial communications.

SW6

SW7

BAUD

OFF

OFF

300

OFF

ON

1200

ON

OFF

2400

ON

ON

9600

Table 5.1: Baud Rate Configuration Table.

Make sure that the baud rate on your computer is set to the same baud rate.

The STM-100 / MF is shipped from the factory set at 9600 baud. To set an IBM-
PC to the same baud rate, type in:

mode com1:9600,8,n,1 <Enter>

This will set the IBM for 9600 baud, 8 data bits, no parity, and 1 stop bit on its

COM1 port. It is important to use 8 data bits because the communication
protocols use all 8 of them.