Table 2.1: baud rate switch settings – INFICON STM-100/MF Thin Film Deposition Monitor User Manual

Page 30

Advertising
background image

SW6

SW7

BAUD

OFF

OFF

300

OFF

ON

1200

ON

OFF

2400

ON

ON

9600

Table 2.1: Baud Rate Switch Settings

SW8, SW9, SW10, SW11, SW12 DEVICE ADDRESS SELECTION

These switches in combination select the device address for the
instrument when operating with the SECS RS-232 protocol or with
either the IEEE-488 or BITBUS communications option cards.
These options allow

multiple devices to operate on a single bus,

and only the individual addressing prevents response conflicts.
Switch coding is in binary 1 of 32 format with SW8 the MSB and
SW12 the LSB. These switches are sensed only during power-
up. See Table 2.2 for a list of addresses.

OPERATION AND PROGRAMMING

Page 2 - 10