Command list for stm-100 / mf, Table 5.2: communication command summary, Detailed command description – INFICON STM-100/MF Thin Film Deposition Monitor User Manual

Page 63: Section 5.6, Software version code, Shutter relay control, Zero thickness and timer, Computer interfacing page 5 - 8, Parameter example comment

Advertising
background image

Parameter

Example

Comment

Sent

Returned

@

None

@

STM100C5

Firmware Version and inst. model number

A

[@,!,?]

A@

Turn shutter relay off

B

None

B

Zeros timer and thickness

C

None

C

Zeros thickness

D

None

D

Zeros timer

E

=[0.500 - 99.99],[?]

E=1.23

Set current film density to 1.23

F

=[0.100 - 9.999],[?]

F?

1.234

Returns current film Z factor

G

=[0 - 9999999],[?]

G=550

Set current film end thickness to 550Å

H

=[0 - 9999999],[?]

H=10560

Set current film setpoint thickness to 10.56KÅ

I

=[00:00 - 99:59],[?]

I=15:30

Set current film setpoint timer to 15:30

J

=[10.0 - 399],[?]

J = 80.1

Set current film tooling to 80.1%

K

[@,!,?]

K?

@

Return test mode status, was off

L

None

L

Acknowledge "a" response

M

None

M

!

Return crystal fail status, was failed

O

None

O

!

Return setpoint timer relay, was closed

P

None

P

@

Return end thickness relay, was open

Q

None

Q

C

Return remote inputs, ZERO TIMER,

ZERO THICK active

R

None

R

193

Return config switch, switches 5,6,12 on

S

None

S

-0001595

Return thickness value -1.595KÅ

T

None

T

(sp)012.4

Return rate, 12.4 Å/s

U

None

U

5319234

Return sensor freq., 5,319,234Hz

V

None

V

012.4

Return crystal life, 12.4% left

W

None

W

+12:45

Return timer, 12 min 45 sec counting up

X

None

X

(sp)0000201

Return log thickness, 201 Å

Y

None

Y

+49:21

Return log timer, since THICK ZERO.

Z

None

Z

-012.3

Return log rate, -12.3 Å/sec

a

None

a

A

Return RESET Status, true

b

None

b

Set parameters to default values

c

[@,!,?]

c?

!

Query beeper status, returned active

d

None

Internal use commands

e

Mask

Internal use commands

f

None

Internal use commands

g

None

Internal use commands

h

None

Internal use commands

i

[1-9,?]

i5

Sets the current film to number 5.

j

[1-9],[0.500 - 99.99]|[?]

Sets / queries the selected film density.

k

[1-9],[0.100 - 99.99]|[?]

Sets / queries the selected film z-factor.

l

[1-9],[0 - 9999000]|[?]

Sets / queries the selected film End Thickness

m

[1-9],[0 - 9999000]|[?]

Sets / queries the selected film

set point relay thickness

n

[1-9],[00:00 - 99:59]|[?]

Sets / queries the selected film timer relay

o

[1-9],[10.0 - 399]|[?]

Sets / queris selected film tooling factor

Table 5.2: Communication Command Summary.

DETAILED COMMAND DESCRIPTION

_________________________________________________________________________

Command:

@

Parameter:

None

Description:

Returns a string indicating the product model number and the software revision level.
The form is "STM100XY" where X is major (A.B,..) and Y is minor digits (0..9).

_________________________________________________________________________

Command:

A

Parameter:

[@,!,?]

Description:

Turns the shutter relay on and off. A "@" will turn the shutter off and a "!" will turn it
on. The indicator on the LCD will indicate the new state. A "?" will send back "@" or
"!" as the current status.

_________________________________________________________________________

Command:

B

Parameter:

None

Description:

Zeros the timer and thickness. This duplicates the ZERO button on the front panel.

Command List For