I/O INTERFACE CONNECTION

This connector provides the user interface to the 4 sets of relay output

contacts as well as the 4 remote input command lines.


RELAY CONTACT RATINGS

2.5 AMPS - 120VAC MAX. - 120VA

Normally Open - SPST CONTACTS

I/O CONNECTOR DEFINITIONS
TYPE

15 Pin "D" Male

MATING CONNECTOR

AMPHENOL 117-DA15S OR EQUIVALENT

FUNCTION

PIN NUMBER

Crystal Fail

1, 2

Elapsed Timer

3, 4

Shutter Relay

5, 6

Thickness Setpoint 7, 8

Ground

13, 14, 15

Shutter Open

12

Shutter Close

11

Zero Thickness

10

Zero Timer

9

Ground

13, 14, 15

RELAY OUTPUTS

The STM-100 / MF is standard equipped with four SPST 2.5 AMP relays

rated to 120VAC and to 120VA. These contacts are normally open and become
active when their respective function is true. Continuous display of the status of
each relay is provided by legends on the LCD display. The available relay
functions are:

1. CRYSTAL FAIL - This relay provides a contact closure whenever the sensor

crystal fails to operate properly or has exceeded its operating range.
The LCD legend for this relay status is (XTAL FAIL). The legend will
also flash as a warning to the user.

2. ELAPSED TIMER - This relay provides a contact closure when the elapsed

timer has counted down to a value of zero from a setpoint value other
than zero. A setpoint value of zero disables this relay function. The
display will continue to count in an upward fashion after a count of
zero is reached. The relay contacts will remain closed until the front
panel ZERO button or a remote TIMER ZERO event occurs, or a new
timer number is entered. The LCD legend for this relay is (SP TIME).

SECTION 3.4

I/O Connector

RELAYS

INPUTS

Relay Outputs

Crystal Fail

Elapsed Timer

Page 3 - 3