INFICON VGC401 Single-Channel Controller User Manual

Page 10

Advertising
background image

10

tinb01e1-f (2011-07) VGC401.om

Number 2

(parallel)

Caution

Do not connect more than one
gauge at the same time.

SENSOR connector

15-pin D-Sub, female
RJ45 (FCC68), female
(pin assignment

→  23)


Front panel
HOST (remote control)

via 3 keys
via RS232C interface


Measurement ranges

depending on gauge
(

→ [1] … [21])

Measurement error

gain error
offset error


≤0.02% FSr
≤0.05% FSr

Measurement rate

analog
digital


100 / s
50 / s

(BPG, HPG, BCG,

CDGxxxD

1)

)

10 / s

(BAG)

Display rate

10 / s

Filter time constant

slow
normal (nor)
fast


750 ms (f

g

= 0.2 Hz)

150 ms (f

g

= 1 Hz)

20 ms (f

g

= 8 Hz)

Pressure units

mbar, Pa, Torr, Micron

Zero adjust

for linear gauges

Correction factor

for logarithmic gauges
0.10 … 10.00

A/D converters

resolution >0.001% FSr
(The measurement values of
BPG, HPG, BCG, BAG and
CDGxxxD are transmitted
digitally.)

1)

CDG025D, CDG045D, CDG100D, CDG160D