INFICON VGC401 Single-Channel Controller User Manual
Page 39

tinb01e1-f (2011-07) VGC401.om
39
Available for:
Pirani
(PSG)
Pirani/Capacitive
(PCG)
Cold cathode
(PEG)
Cold cathode/Pirani
(MPG)
Hot cathode
(BAG)
Hot cathode/Pirani
(BPG, HPG)
;
Capacitive
(CDG)
Hot cathode/Pirani/Capacitive
(BCG)
When the offset correction is activated, the stored offset
value is subtracted from the actual measurement value.
This allows measuring relative to a reference pressure.
The offset correction affects:
;
the displayed measurement value
the displayed threshold value of the switching
functions
the analog output at the CONTROL connector
(
→
23)
Value
Ö
Offset correction
deactivated
e.g.:
Ö
Offset correction
activated
para
Ö
Press briefly:
Activate offset correction.
para
Ö
Press key >2 s:
The offset value is read-
justed. The actual measure-
ment value is accepted as
new offset value.
para
Ö
Press briefly:
Deactivate offset correction.