69

Transmit:

BAU

[,x] <CR>[<LF>]


└─ x = 0 –> 9600 baud (default)

1 –> 19200 baud

2 –> 38400 baud

As soon as the new baud rate has been entered,

the report signal is transmitted at the new
transmission rate.

Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x

<CR><LF>


└─ Transmission rate


Transmit:

EUM

[,x] <CR>[<LF>]


└─ x = 0 –> Manually

1 –> Automatically

(default)

Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x

<CR><LF>


Transmit:

FUM

[,x] <CR>[<LF>]


└─ x = 0 –> Automatically

(default)

1 –> Filament 1

2 –> Filament 2

Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x

<CR><LF>


Transmit:

SAV

[,x] <CR>[<LF>]


└─ x = 0 –> Save default

parameters

1 –> Save user parameters

Receive: <ACK><CR><LF>

Transmission rate

Emission

Filament

Save parameters to
EEPROM