INFICON VGC401 Single-Channel Controller User Manual
Page 69
tinb01e1-f (2011-07) VGC401.om
69
Transmit:
BAU
[,x] <CR>[<LF>]
│
└─ x = 0 –> 9600 baud (default)
1 –> 19200 baud
2 –> 38400 baud
As soon as the new baud rate has been entered,
the report signal is transmitted at the new
transmission rate.
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x
<CR><LF>
│
└─ Transmission rate
Transmit:
EUM
[,x] <CR>[<LF>]
│
└─ x = 0 –> Manually
1 –> Automatically
(default)
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x
<CR><LF>
Transmit:
FUM
[,x] <CR>[<LF>]
│
└─ x = 0 –> Automatically
(default)
1 –> Filament 1
2 –> Filament 2
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x
<CR><LF>
Transmit:
SAV
[,x] <CR>[<LF>]
│
└─ x = 0 –> Save default
parameters
1 –> Save user parameters
Receive: <ACK><CR><LF>
Transmission rate
Emission
Filament
Save parameters to
EEPROM