1 measurement mode – INFICON VGC401 Single-Channel Controller User Manual
Page 60
Advertising
60
tinb01e1-f (2011-07) VGC401.om
Transmit:
PR1
<CR>[<LF
>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x,sx.xxxxEsxx
<CR><LF>
│ │
│ └─ Measurement value
1)
│ [in current pressure unit]
│
└─ Status, x =
0 –> Measurement data okay
1 –> Underrange
2 –> Overrange
3 –> Sensor error
4 –> Sensor off (BAG, PEG)
5 –> No sensor
6 –> Identification error
7 –> Error BAG, BPG, HPG, BCG
1)
The 3
rd
and 4
th
decimal are always 0, except for
the CDG gauge.
5.2.1 Measurement
Mode
Measurement data
Advertising