Transmit:

ERR

<CR>[<LF>]

Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: xxxx

<CR><LF>


└─ x =

0000 –> No error

1000 –> Controller error

(See display on front panel)

0100 –> NO, HWR No hardware

0010 –> PAR, Inadmissible parameter

0001 –> SYN, Syntax error

The ERROR word is cancelled when read out. If
the error persists, it is immediately set again.


Transmit:

RES

[,x] <CR>[<LF>]


└─ x = 1 –> Reset

Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: [x]x,[x]x,...

<CR><LF>


└─ List of all present error messages

xx =

0 –> No error

1 –> Watchdog has responded

2 –> Task fail error

5 –> EPROM error

6 –> RAM error

7 –> EEPROM error

9 –> DISPLAY error

10 –> A/D converter error

11 –> Sensor error (e.g. filament

rupture, no supply)

12 –> Sensor identification error



Error status