INFICON VGC401 Single-Channel Controller User Manual
Page 63

tinb01e1-f (2011-07) VGC401.om
63
Transmit:
ERR
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: xxxx
<CR><LF>
│
└─ x =
0000 –> No error
1000 –> Controller error
(See display on front panel)
0100 –> NO, HWR No hardware
0010 –> PAR, Inadmissible parameter
0001 –> SYN, Syntax error
The ERROR word is cancelled when read out. If
the error persists, it is immediately set again.
Transmit:
RES
[,x] <CR>[<LF>]
│
└─ x = 1 –> Reset
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: [x]x,[x]x,...
<CR><LF>
│
└─ List of all present error messages
xx =
0 –> No error
1 –> Watchdog has responded
2 –> Task fail error
5 –> EPROM error
6 –> RAM error
7 –> EEPROM error
9 –> DISPLAY error
10 –> A/D converter error
11 –> Sensor error (e.g. filament
rupture, no supply)
12 –> Sensor identification error
Error status
Reset