Ö gauge dependent – INFICON VGC401 Single-Channel Controller User Manual

Page 36

Advertising
background image

36

tinb01e1-f (2011-07) VGC401.om

lowe

r

thres

hold

limit

uppe

r

thres

hold

limit

all values in mbar, Cor = 1

1.5Ч10

3

5Ч10

2

1Ч10

-2

1Ч10

3

2Ч10

-3 *)

FSr

FSr / 1000

1Ч10

-9

1Ч10

-8

1Ч10

-10

1Ч10

-1

1Ч10

3

1Ч10

-6

1.5Ч10

3

2Ч10

-3 *)

1Ч10

-8

1Ч10

3

1Ч10

-8

5Ч10

-9

1Ч10

3

FSr

FSr / 1000

*)

2×10

-4

mbar, if PrE is activated (

→  46)

The minimum hysteresis between the upper and
lower switching threshold is at least 10% of the
lower threshold or 1% of the set full scale value.
If the value of the minimum hysteresis drops be-
low these values, the upper threshold is auto-
matically adjusted. This prevents unstable
states.

Value

The upper switching threshold
(Setpoint high) defines the pres-
sure at which the switching func-
tion is deactivated when the pres-
sure is rising.

e.g.:

Ö

gauge dependent

(

table).

If another gauge type is con-
nected, the VGC401 auto-
matically adjusts the threshold
if required.

Advertising