INFICON VGC401 Single-Channel Controller User Manual
Page 73
Advertising
tinb01e1-f (2011-07) VGC401.om
73
Transmit:
TDI
[,x] <CR>[<LF>]
│
└─ x = 0 –> Stops the test – display
according to current
operating mode
(default)
1 –> Starts the test – all
LEDs
on
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x
<CR><LF>
│
└─ Display test status
Transmit:
TAD
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive:
[x]x.xxxx, x.xxxx, x.xxxx <CR><LF>
│ │ │
│ │ └─ ADC channel 2
│ │
Gauge
│ │
identification
│ │
[0.0000 …
│ │ 5.0000
V]
│ │
│ └─ ADC channel 1
│ Measurement signal
│ (negative portion)
│ [0.0000 … 5.0000 V]
│
└─ ADC channel 0
Measurement signal (positive
portion) [0.0000 … 11.0000 V]
Display test
ADC test
Advertising