tinb01e1-f (2011-07) VGC401.om

75


"Transmit (T)" and "Receive (R)" are related to the host.

T:

TID

<CR> [<LF>]

R: <ACK> <CR> <LF>
T: <ENQ>
R: PSG <CR> <LF>

Request for gauge identification
Positive acknowledgement
Request for data transmission
Gauge identification

T:

SP1

<CR> [<LF>]


R: <ACK> <CR> <LF>
T: <ENQ>
R: 1.0000E-09,9.0000E-07 <CR> <LF>

Request for parameters of
switching function (setpoint)
Positive acknowledgement
Request for data transmission
Thresholds

T:

SP1

,6.80E-3,9.80E-3 <CR> [<LF>]


R: <ACK> <CR> <LF>

Modification of threshold values of
switching function (setpoint)
Positive acknowledgement

T:

FOL

,2 <CR> [<LF>]


R: <NAK> <CR> <LF>
T: <ENQ>
R: 0001 <CR> <LF>
T:

FIL

,2 <CR> [<LF>]

R: <ACK> <CR> <LF>
T: <ENQ>
R: 2 <CR> <LF>

Modification of filter time constant
(syntax error)
Negative acknowledgement
Request for data transmission
ERROR word
Modification of filter time constant
Positive acknowledgement
Request for data transmission
Filter time constant

T:

PR1

<CR> [<LF>]

R: <ACK> <CR> <LF>
T: <ENQ>
R: 0,8.3400E-03 <CR> <LF>
T: <ENQ>
R: 1,8.0000E-04 <CR> <LF>

Request for measurement data
Positive acknowledgement
Request for data transmission
Status and pressure
Request for data transmission
Status and pressure