INFICON VGC401 Single-Channel Controller User Manual

Page 38

Advertising
background image

38

tinb01e1-f (2011-07) VGC401.om

The full scale value of the measurement range (Full
S

cale range) of the linear gauges has to be defined by

the user; the full scale value of logarithmic gauges is
automatically recognized.
Available for:

†

Pirani

(PSG)

†

Pirani/Capacitive

(PCG)

†

Cold cathode

(PEG)

†

Cold cathode/Pirani

(MPG)

†

Hot cathode

(BAG)

†

Hot cathode/Pirani

(BPG, HPG)

;

Capacitive

(CDG)

†

Hot cathode/Pirani/Capacitive

(BCG)



Value

e.g.:

Ö

0.01 mbar
0.01 Torr, 0.02 Torr, 0.05 Torr
0.10 mbar, 0.25 mbar, 0.50 mbar
0.10 Torr, 0.25 Torr, 0.50 Torr
1 mbar, 2 mbar, 5 mbar
1 Torr, 2 Torr, 5 Torr
10 mbar, 20 mbar, 50 mbar
10 Torr, 20 Torr, 50 Torr
100 mbar, 200 mbar, 500 mbar
100 Torr, 200 Torr, 500 Torr
1000 mbar, 1100 mbar
1000 Torr

2 bar, 5 bar, 10 bar, 50 bar

Conversion table

Appendix,  81


For displaying the offset correction and zero adjustment
of the gauge and adjustment to the currently measured
value (in the range -5 ... +110% of the full scale setting).

Measurement range
of capacitive gauges

Offset correction of
the controller