2 parameter mode – INFICON VGC401 Single-Channel Controller User Manual

Page 64

Advertising
background image

64

tinb01e1-f (2011-07) VGC401.om


Transmit:

DGS

[,x] <CR>[<LF>]


└─ x = 0 –> off (default)

1 –> on (3 min.)

Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x

<CR><LF>


└─ Degas status



Transmit:

SP1

[,x.xxEsx,x.xxEsx] <CR>[<LF>]

│ │
│ └─ Upper threshold

1)

│ [in current pressure
│ unit]
│ (default = depending
│ on gauge)


└─ Lower threshold

1)

[in current pressure unit]
(default = depending on
gauge)

1)

Values can be entered in any format. They are internally

converted into the floating point format.


Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x.xxxxEsxx,x.xxxxEsxx

<CR><LF>

│ │
│ └─ Upper threshold
│ [in current pressure unit]


└─ Lower threshold

[in current pressure unit]



5.2.2 Parameter

Mode

Degas

Threshold value
setting, allocation