Connect the peripheral components to the

CONTROL connector on the rear of the unit.
Use a screened cable (electromagnetic com-
patibility).

DANGER

DANGER: protective low voltage
According to EN 61010, voltages ex-
ceeding 30 VAC or 60 VDC are haz-
ardous.
Only connect a protective low voltage
(SELV).

Pin assignment of
the male 9-pin D-Sub
appliance connector:

5

1

9

6

Pin Signal

1
7
5

Analog output

0 … +10 VDC

Chassis = GND
HV_H on +24

V

off

0

V

The control over this signal is placed superior to the key
operation.

4
3
2

Pressure below
threshold

Pressure above
threshold or power
supply turned off

Error signal

9
8

No error

Error or power
supply turned off

Supply for relays with higher switching power

6
7

+24 VDC, 200 mA
Chassis = GND

Fuse-protected at 300 mA with PTC
element, self-resetting after power
off or pulling the CONTROL con-
nector. Meets the requirements of a
grounded protective extra low
voltage (SELV).

The analog output (pin 1) differ from the

displayed value by no more than ±50 mV.


Pin assignment
Contact positions
CONTROL