INFICON VGC401 Single-Channel Controller User Manual

Page 68

Advertising
background image

68

tinb01e1-f (2011-07) VGC401.om

Transmit:

COR

[,[x]x.xxx] <CR>[<LF>]


└─ 0.100 … 10.000

(default = 1.000)

Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: [x]x.xxx

<CR><LF>


└─ Correction factor


Transmit:

DCD

[,x] <CR>[<LF>]


└─ x = 2 –> 2 digits (default)

3 –> 3 digits

Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x

<CR><LF>


└─

Number of digits


When the PrE (

→  46) is ON and the pressure is in the

range p<1.0E-4 mbar the display resolution of the PSG
and PCG Gauges is reduced by one decimal digit.


Transmit:

FIL

[,x] <CR>[<LF>]


└─ x = 0 –> fast

1 –> medium (default)

2 –> slow

Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x

<CR><LF>


└─ Filter time constant


Correction factor

Number of digits in
the display

Measurement value
filter

Advertising