INFICON VGC401 Single-Channel Controller User Manual
Page 68
68
tinb01e1-f (2011-07) VGC401.om
Transmit:
COR
[,[x]x.xxx] <CR>[<LF>]
│
└─ 0.100 … 10.000
(default = 1.000)
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: [x]x.xxx
<CR><LF>
│
└─ Correction factor
Transmit:
DCD
[,x] <CR>[<LF>]
│
└─ x = 2 –> 2 digits (default)
3 –> 3 digits
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x
<CR><LF>
│
└─
Number of digits
When the PrE (
→ 46) is ON and the pressure is in the
range p<1.0E-4 mbar the display resolution of the PSG
and PCG Gauges is reduced by one decimal digit.
Transmit:
FIL
[,x] <CR>[<LF>]
│
└─ x = 0 –> fast
1 –> medium (default)
2 –> slow
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x
<CR><LF>
│
└─ Filter time constant
Correction factor
Number of digits in
the display
Measurement value
filter